Product Information

ADMA High Sensitive - ELISA

Catalog-No.

EA209/96
Highly sensitive Enzyme Immuno Assay
for the quantitative determination of
Asymmetric Dimethylarginine (ADMA) in Serum or Plasma of Mice, Rats and in Cell Culture Media

Format

12x8 Determinations

Procedure

see instruction for use

Clinical Fields

Diseases

Normal Value

Specificity

Substance   ED-50-Value (μmol/l)   Cross Reactivity (%)
ADMA          0.745                          100
Arginin         3.993                          < 0.020
SDMA          2.990                             0.025
NMMA        51                                   1.460

Sensitivity

Sensitivity: 0,01 µmol/l

Characteristics

Lower limit of detection: 0.01 µmol/l

Suitable for small sample volumes (5-20 µl).

Samples can be run directly in the assay without any filtration

High precision is guaranteed by over night incubation

Detailed testing in RPMI and DMEM cell culture medium

No cooling procedure during the transportation

No cross reactivity with L-Arginine, SDMA and N-Monomethylarginine

No cross reactivity with drugs

Excellent correlation with LC-MS/MS analysis

Kits with long expiry date (up to 21 month) are in stock ready for shipment
 

Instruction Sheet